Over 100 Washington Auctions End Today - Bid Now

spacer
Listing 2898291 Listing 2898291 Listing 2898291 Listing 2898291 Listing 2898291 Listing 2898291
Listing 2898291 Listing 2898291 Listing 2898291 Listing 2898291 Listing 2898291 Listing 2898291

Wanted ALCON INFINITI Phacoemulsifier

  • Condition : Used - Good
  • Qty. Available : 1
  • Date : May 09, 2024
  • Listing # : 2898291
Advertisement

Item Description

Seller assumes all responsibility for this listing..

Listing: #2898291

  • Condition: Used - Good
  • Brand: ALCON
  • Type: Phacoemulsifier
  • Model: INFINITI
  • Software Version: 2.06
Always looking for Alcon Infiniti Phacoemulsifiers in any condition. We will fully PM and make patient ready.

We have the ability to supply parts for the ALCON INFINITI System. We also offer a full service course to the BIOMED so that you can fully maintain this system on your own. If you need parts, support or training, we can do it.

All the systems we sell are fully up to date, and patient ready.

1yr. Warranty on all systems

All variations, software versions, are wanted. We guarantee all systems, Preventive Maintenance, Quality Assure all systems before going out. Please call or email for availability and specifications. AND If you have any phaco system to sell, let us know.

*******************************************************

This list of error code Alcon Infiniti is a partial list. If you need additional information, contact us and we will be happy to assist, just as we have trained alot of biomeds on the Alcon Infiniti Service and the Alcon Infiniti PM procedure. We will be happy to help you with parts and training.

THE LISTED ERROR CODES ARE SHOWN ON THE INFINITI VISION SYSTEM DISPLAY PANEL WHEN THE SYSTEM DETECTS A PROBLEM. THE ERROR CODES ARE SEPARATED BETWEEN ADVISORIES, WARNINGS, AND FAULTS.

Alcon Infiniti Error Code 160: Reinsert FMS. Replace FMS if Problem Persists.

Alcon Infiniti Error Code 161-164: Please Check Fittings and Reprime. Replace FMS if Problem Persists.

Alcon Infiniti Error Code 165-167: Flow Obstruction. Please Check Handpiece Free Flow

Alcon Infiniti Error Code 169: Irrigation Pressure is Low. Please Check Bottle and Fittings.

Alcon Infiniti Error Code 170: Reflux Terminated. Reflux Fluid Volume Depleted.

Alcon Infiniti Error Code 171: Excessive Pressure in Drain Bag. Replace FMS.

Alcon Infiniti Error Code 175-176: Infusion Pressure Sensor Calibration Error. Remove FMS. If Problem Persists, Contact Alcon Technical Services.

Alcon Infiniti Error Code 180: Invalid FMS ID. Replace FMS.

Alcon Infiniti Error Code 181: Excessive Ambient Light. Unable to Calibrate FMS ID Sensors.

Alcon Infiniti Error Code 182: Excessive Ambient Light. Unable to Read FMS ID.

Alcon Infiniti Error Code 183: Remove MultiPak FMS, See MultiPak Directions for Use

Alcon Infiniti Error Code 250: Tuning in Air

Alcon Infiniti Error Code 251: Insert Handpiece

Alcon Infiniti Error Code 252: Multiple Handpieces Detected. Remove a Handpiece.

Alcon Infiniti Error Code 254: Loose Tip

Alcon Infiniti Error Code 256-270: Replace Handpiece. If Problem Persists After Restarting System, Contact Alcon Technical Services.

Alcon Infiniti Error Code 271: Two Handpieces Detected. Remove a Handpiece.

Alcon Infiniti Error Code 272-274: Replace Handpiece. If Problem Persists After Restarting System, Contact Alcon Technical Services.

Alcon Infiniti Error Code 276: Ultrasound Error. Release Treadle and Retry. If Problem Persists After Restarting System, Contact Alcon Technical Services.

Alcon Infiniti Error Code 277: Handpiece Disconnected While Applying U/S Power. Release Treadle, then Insert and Tune Handpiece.

Alcon Infiniti Error Code 278: Replace Handpiece. If Problem Persists After Restarting System, Contact Alcon Technical Services.

Alcon Infiniti Error Code 279: Unknown Handpiece Detected.

Alcon Infiniti Error Code 280: Unsupported Handpiece Detected.

Alcon Infiniti Error Code 281: Cautery Compliance Error. If Problem Persists After Restarting System, Contact Alcon Technical Services.

Alcon Infiniti Error Code 282: Coagulator Error. If Problem Persists After Restarting System, Contact Alcon Technical Services.

Alcon Infiniti Error Code 350: Footswitch Failure Detected. Check Footswitch, Clean Under Rear Section of Treadle, and Remove Debris if Present. (Reference Maintenance Section of Operator’s Manual.) Ensure Treadle is Not Depressed, then Reset Footswitch. If Condition Persists, Contact Alcon Technical Services.

Alcon Infiniti Error Code 351: Footswitch Failure Detected. Check and Reset Footswitch. If Condition Persists, Contact Alcon Technical Services.

Alcon Infiniti Error Code 352: Footswitch Failure, Replace Footswitch.

Alcon Infiniti Error Code 353: Footswitch Failure Detected, Check and Reset Footswitch. If Condition Persists, Contact Alcon Technical Services.

Alcon Infiniti Error Code 450: Footswitch is Depressed, Release Footswitch Before Pressing Prime FMS, Fill, or Test Handpiece.

Alcon Infiniti Error Code 451: Can Not Recognize Footswitch. Please Check Footswitch Connection and Reset Footswitch. If Condition Persists, Contact Alcon Technical Services.

Alcon Infiniti Error Code 460: Please Install Footswitch.

Alcon Infiniti Error Code 461: Infiniti ® Backup Power Service Needed, Contact Alcon Technical Services.

Alcon Infiniti Error Code 463: Invalid Language(s) Found During Initialization. One or More Installed Languages May Not be Available.

Alcon Infiniti Error Code 464: The Language Specified by System Settings is Invalid.

Alcon Infiniti Error Code 465: The Tune Sequence was Interrupted by removal or the Handpiece.